Thursday, 15 February 2018

SOP for Use, Calibration and maintenance of Micropipettes.


1.0         Objective:
               The purpose of this SOP is to provide written procedure for Use, Calibration and maintenance of Micropipettes.

2.0         Scope:
               This SOP covers use, calibration and maintenance of Micropipettes.

3.0       Responsibility:  
               Jr. Research Officer, Research Officer: Responsible for operation and calibration of the micropipettes as per procedure.
    Head of Department: Responsible for calibration and maintenance, timely as per schedule.
         QA Officer/QA Manager: Review the records and governing the document.

4.0       Procedure:
Use:
4.1  Attach the tip specific to the required volume to the tip                    cone. 
       
  4.2  Hold the pipette in an upright position while aspirating the liquid. The grippy should rest on the index finger. 
          
4.3      Before beginning the actual pipetting work, fill and empty the tip 2-3 times with the solution that is to be pipetted.

4.4       Depress the push button to the first stop.    

4.5    Dip the tip into the solution and slowly release the push button to the ready position. Withdraw the tip from the liquid, touching it against the edge of the container to remove excess liquid.

4.6    Deliver the liquid by gently pressing the push button to the first stop. After a delay of about one second, continue to press the push button to the second stop. This will empty the tip.
              
4.7     Release the push button to the ready position.

4.8    Release the tip in a waste receptacle by pressing the tip                   ejector with the thumb. 
  
Calibration:

4.9    The following are required for calibration of the micropipettes.

4.10    Calibrated balance with the following readable graduation for various volume range pipettes.

S.No
Volume Range
Readable graduation
1
Less than 100µl
0.01 mg
2
Above 100µl
0.1 mg

               Calibrated thermometer
               Vessel for weighing
               Test Liquid: Distilled water
               Temperature of Test Liquid: 15 to 30 °C
4.11      Place a calibrated thermometer in the test liquid (distilled water) and record the   temperature.

4.12      Pre-wet the tip for 3-5 times with the distilled water.

4.13      Set the pipette delivery volume at minimum range of the                pipette.

4.14     Place the weighing vessel on the pan of the balance and                  tare   the weight.

4.15     Follow the pipetting technique as described in section 4.0.

4.16       Record the measured weight (in mg) and tare the weight on the pan.

4.17     Follow the steps 4.15 to 4.16 for 10 pipettings.

4.18     Set the pipette delivery volume at maximum range of the                pipette.

4.19     Follow the steps 4.15 to 4.16 for 10 pipettings.

4.20     Fixed volume micropipette will be calibrated only at one                  volume.

5.0      Calculations:

5.1    Conversion formula for measured weight in mg, to volume in            microlitres
            V = W/d
Where V: volume in µl
W: weight in mg
d: density of water in g/ml at the calibration temperature.

5.2       Inaccuracy (%) = (mean volume X 100/set volume)-100

5.3       Imprecision (CV %) = 100 X SD/mean volume
CV: Co-efficient of variation
SD: Standard Deviation
Acceptance criteria as per ISO8655


Range Of Prepitte
Volume (µl)
Inaccuracy(%)
Imprecision(CV%)
Should Be less than
0.2-2 µl
0.2
±40.0
20.0
2
±4.0
2.0
0.5-10 µl
1
±12.0
8.0
10
±1.2
0.8
2-20 µl
2
±10.0
5.o
20
±1.0
0.5
5-50 µl
5
±10.0
4.0
50
±1.0
0.4
10-100 µl
10
±8.0
3.0
100
±0.8
0.3
20-200 µl
20
±8.0
3.0
200
±0.8
0.3
100-1000 µl
100
±8.0
3.0
1000
±0.8
0.3
200-1000 µl
200
±4.0
1.5
1000
±0.8
0.3
1-5 ml µl
1000
±4.0
1.5
5000
±0.8
0.3
2-10 ml µl
2000
±3.0
1.5
10000
±0.6
0.3



          
   
  






























 Calibration frequency:
6.1      Calibrate the micropipettes once in three months                              (+ extended 10 days)

6.2      Calibrate the pipette upon receipt from the supplier after                 repair (on receipt).

6.3      Recalibrate the pipette on failure of calibration                                (Recalibration).

7.0                Precautions:

Take the following precautions when using a micropipette:

7.1    NEVER rotate the volume adjuster beyond the upper or                  lower range of the pipette.

7.2        NEVER invert or lay the micropipette down with a filled tip; fluid can run back into the piston.

7.3       NEVER let the plunger snap back after withdrawing or                   expelling fluid; this could damage the piston.

7.4       NEVER immerse the barrel of the micropipette in fluid.

7.5       NEVER flame the tip of the micropipette.

7.6      When not in use, keeping the micropipette in its maximum volume range prolongs the life span of the spring.

8.0                  Routine maintenance:

8.1                Clean the instrument with dry cloth.

8.2         Clean the electrode with clean and dry tissue paper.

9.0         Documentation:

9.1   Annexure – 1 Density (g/ml) of water at different temperature.

9.2 Annexure –2 Micropipette Calibration Format   XXX/FRM/000

10.0       History of Revision:
Revision No.
Effective Date
Revision details
Reason for revision





                                               Annexure 1



























                                                         



                                                                 Annexure 2

MICROPIPETTE CALIBRATION FORMAT

Pipette ID:                                                                      Volume Range:
Date of calibration:                                                        Temperature of distilled water:
Density of water:


S.No.
Minimum delivery volume
Maximum delivery volume
Set Volume (µl)
Measured Weight (mg)
Volume (µl)
Set Volume (µl)
Measured Weight (mg)
Volume (µl)
1






2




3




4




5




6




7




8




9




10




Mean Volume (µl)

Mean Volume (µl)

Inaccuracy (%)

Inaccuracy (%)

Standard Deviation (%)

Standard Deviation (%)

Imprecision (%)

Imprecision (%)


Balance Used (ID)

Thermometer Used (ID)

Result:
Calibrated by
Signature and Date
Calibration due on





SOP .: XXX/SOP/000                                                         Format No.:XXX/FRM/000


No comments:

Post a Comment

requirement of Q.C in Daman

  We have requirement of Q.C Working area : Daman/Dalwada Exp: Fresher Gender: Male Qualification: Any Graduated Interested Person Call On t...