1.0 Objective:
8.0 History
of Revision:
To provide written procedure for
operation of Atomic Absorption Spectrometer.
2.0 Scope:
This SOP covers operation of
Atomic Absorption Spectrometer
3.0 Responsibility:
Jr.
Research Officer, Research Officer: Responsible for operation, calibration and
maintenance of the instrument as per procedure.
Head of Department: Responsible
for calibration and maintenance, timely as per schedule.
QA Officer/QA Manager: Review the records and governing the document.
4.0 Procedure:
4.1 Ensure that the instrument is clean
and free from dust.
4.2 Operating Procedure:
4.2.1 Open the outlet of the acetylene cylinder
supply.
4.2.2 Switch ‘ON’ the air compressor. The
pressure must be 13-15 psi.
4.2.3 Switch ‘ON’ the computer system, monitor
and the instrument.
4.2.4 Double click on the spectrAA icon or
click on start menu, select programs and then select Varian AAS and SpectrAA.
4.2.5 After click on Worksheet icon load
worksheet dialogue. Open new worksheet then give the name of worksheet and
click ok.
4.2.6 Click on icon add methods and select the
method. Place a Hollow Cathode Lamp as per selected method in a lamp holder
positioned to the right of the instrument.
4.2.7 Click
on icon Edit methods and open the dialogue box.
4.2.8 Don’t
use SIPS then Sampling Mode Manual, Measurement Mode, Integration Mode, Fill
standard concentration and click OK.
4.2.9 Click
Tab analysis and click on icon Optimize, select method and click OK. Then flame
optimization windows open. Align the lamp up to maximum signal and click on
optimize signals and click Inst. Zero then click OK.
4.3 Burner System:
4.3.1 Open the Burner door and confirm the
standard acetylene burner head is installed.
Close the burner door.
4.3.2 Setting the burner head (by turning the
height control knob) lowers and upper side so that it is clear of the sample
beam. Click on instrument zero. Slowly turn the burner height control knob
counter until the absorbance display read zero again.
4.3.3 Very
slowly turn the burner height control knob until the display read zero again.
4.4 Lighting
The Flame:
4.4.1 Switch
on the exhaust fan.
4.4.2 Push
the flame ignite button located the front side of the instrument.
4.4.3 Ensure
that the nebulizer uptake tube is placed into a blank solution or purified
water whenever the flame is align.
4.5 Optimizing
Instrument Performance:
4.5.1 Allow the burner system to warm up for 2
minutes.
4.5.2 Aspirate the blank solution and press
Inst. Zero button. Aspirate the known concentration of solution and locate the
burner horizontal and vertical position for maximum absorbance reading.
4.5.3 Adjust the nebulizer for best
sensitivity by firstly turning the nebulizer.
4.6 Analysis of Samples:
4.6.1 To open the main window of analysis,
aspirate the Blank solution.
4.6.2 Click on icon start then asks prepare
the instrument Zero click OK.
4.6.3 After click OK ask Present Solution CAL
ZERO click OK.
4.6.4 After complete the blank ask for
Standards, aspirate the standards then calibration curve is completed.
4.6.5 After complete the Calibration curve
asks for sample aspire. After Sample aspire save the worksheet.
4.7 Procedure to Operate the VGA-76
Accessory:
4.7.1 VGA-76 Accessory attached the front side
of AAS.
4.7.2 Check
the current power supply. Connect the nitrogen cylinder and release pressure about 4 Lb/in2.
4.7.3 Connect the tubing with pump. First all 3
tubes wash with purified water for 20 – 30 minutes.
4.7.4 On the burner Head attached the cell
stand and put the cell.
4.7.5 Connect the cell with VGA-76 with rubber
tube.
4.7.6 Before Analysis the one of the last two
tubes put in Acid Container and other tube put in Reductant Container. Third
tube put on the blank solution.
4.7.7 To start the analysis the sample tube put
in the blank, after the Blank the tube put in standard, After the standard
Calibration Curve Completed the tube put in sample.
4.7.8 After the analysis the sample, tube put
in the water for washing.
4.8 Shutting Down the VGA-76 Accessory:
4.8.1 After the analysis all tubes put in water
for washing. Continue the washing for 20-30 minutes. After washing the tubes
disconnect from the pump.
4.8.2 Close the Nitrogen Cylinder and disconnect
it. The VGA-76 Accessory Remove from AAS.
4.9 Shut Down Procedure:
4.9.1 With
the flame still burning, aspirate the correct rinsing solutions in the sequence
listed below to rinse the nebulizer and burner.
4.9.2 If used aqueous solutions during the
analysis aspirate purified water for 2 minutes.
4.9.3 Push Flame off button located on front
side of the Instrument.
4.9.4 Turn off the gas supplies to the
Spectrometer.
4.10 Report:
4.10.1 For report double click on the spectrAA
icon, click on icon reports. Open the window of
Worksheet
then click next, select solution data, report contents, calibration graphs etc.
4.10.2 Then click next then report is generated,
to see the report click on print preview or print out the report click print
report.
4.11 After completion
exit the soft ware and switch off the instrument.
4.12 Record the usage of instrument in
instrument usage log book (Annexure - 1).
5.0
Precaution:
5.1 COMPRESSED GASES SHOULD ALWAYS BE
HANDLED WITH CARE. To avoid leaks, periodically check the lines and fittings. Use soap water to
check leaks.
5.2 Acetylene, in its free state under pressure may decompose with
explosive violence,therefore NEVER EXCEED A PRESSURE OF 100 kPa [ 15 psi].
5.3 NEVER USE OXYGEN OR OXYGEN-ENRICHED
AIR, AS AN INSTANT FLASHBACK WILL
RESULT.
5.4 Do not touch the quartz lamp envelope
with bare hands. Skin oils may be adsorbed into the quartz. Always use clean
cotton gloves when handling lamps. If contact is made with the lamp surface, clean
immediately with propanol or equivalent.
5.5 The liquid trap MUST be filled with
the same solvent as that used for the analysis.
5.6 Always
allow a burner to cool before removal.
5.7 Do not manually move the arm of the
flame sampler as this may damage the drive mechanisms.
5.8 Do not leave uncovered containers of
volatile organic solvents near the naked flame.
5.9 Do not store solvent in the vicinity
of the atomic absorption spectrometer.
5.10 Ventilate the working area by using an
efficient extraction system to remove burnt gases and toxic compounds that may
be formed.
6.0
Routine maintenance:
6.1
Clean the instrument with dry
cloth
6.2
Clean the burner with burner
cleaning card.
6.3
Check the hoses and plumbing of
gases for leaks.
7.0
Documentation:
7.1 Annexure – 1 Instrument Usage logbook. XXX/FRM/000
Revision No.
|
Effective
Date
|
Revision
details
|
Reason for
revision
|
Annexure - 1
|
|||||
Date
|
Name of Sample
|
A. R. No.
|
Batch No.
|
Analysed by
|
Remarks
|
SOP No. : XXX/SOP/000-00 Format No. : XXX/FRM/000-00
|
No comments:
Post a Comment