1.0 Objective:
To provide written procedure for
operation and calibration of Antibiotic Zone Reader.
2.0 Scope:
This SOP covers the operation and
calibration of Antibiotic Zone Reader.
3.0 Responsibility:
Microbiologist: Responsible for
operation, calibration and maintenance of the instrument. Maintain the required
record.
QA Officer/QAM: Review the
records and governing the document.
4.0 Procedure:
4.1 Ensure that the instrument is visibly clean and free from dust.
4.2 Switch ‘ON’ the
mains switch.
4.3 Turn the left side drum to its maximum
clockwise position, so that Zero Marking on the scale of the right side drum
coincides with the Zero marking on the Vernier scale.
4.4 Keep the Petri-dish on the Aluminum
plate and rotate it manually; so that one end of the Zone image touches the Mark Line
on the Prism.
4.5 Now turn the left-side drum slowly, in
anti-clockwise position, till the other end of the Zone image touches the Mark Line
on the Prism.
4.6 Note the reading directly on the Right
side drum scale. The scale is marked from 0 to 35 mm with each division of 0.2
mm and the Vernier scale is marked with each division of 0.02 mm.
5.0 Calibration:
5.1 Operate
the instrument as per above procedure.
5.2 Put
the standard coin in the appropriate place on antibiotic zone reader.
5.3 Measure
the zone of standard coin and compare with the standard coin result and
observed result.
5.4 Record the result in the calibration
format
6.0
Routine maintenance:
6.1
Clean the instrument with dry
cloth.
6.2
Clean glass & lance with
dry cloth or tissue paper.