1.0 Objective:
The objective of this Standard
Operating Procedure is to lay down the procedure for operating the BOD
Incubator.
2.0 Scope:
This SOP covers operating
procedure of BOD Incubator .
3.0 Responsibility:
Microbiologist:
Responsible for operation of instrument as per procedure and maintain the
record.
QA Officer/QAM: Review the
records and governing the document.
4.0 Procedure:
4.1. Switch ON the mains and instrument; observe
that the LED of Temperature indicator displays the ambient value.
4.2. Press ‘SET’ key (Red Button) to pre set the
desired performance parameter.
4.3. The LED shall display the last setting. Turn
the setting knob clockwise for decrement until the LED displays the desired set
value. Use the second knob for fine setting if required.
4.4. Depress (release) the SET key, The LED shall
regain to display the ambient value of the chamber.
4.5. Now the unit will automatically perform to
attain the set temperature with the on and off of the heater.
4.6. The heater selection switch is provided
having positioned OFF, HIGH and LOW.
4.7. Select the switch initially at HIGH to
attain the set point quickly. Turn it to LOW soon the display is reading near
to the set point for fine and accurate control, thus to prevent the shooting up
the temperature from the set value.
4.8 Clean
the incubator with 5 % Dettol solution or 70 % Isopropyl alcohol.
5.0 Precaution:
5.1 Do not leave the door open for an
extended period.
5.2 Load
the samples quickly but gently and close the door properly with the locking the
front lock.
5.3 Do not overload the sample more then
the chamber volume.
5.4 Avoid the placement of sample so close
to each other so as to allow the proper circulation of heated air for
maintaining even condition through out the chamber.
6.0 Routine maintenance:
6.1 Clean instrument with isopropyl
alcohol. During cleaning, the instrument should be off.
6.2 Wipe the instrument with clean and dry
cloth.
7.0
Documentation:
7.1 Annexure - 1 Temperature record logbook XXX/FRM/000-00
7.2
Annexure - 2 Incubator
logbook XXX/FRM/000-00
8.0 History of Revision:
Revision No.
|
Effective
Date
|
Revision
details
|
Reason for
revision
|
Annexure – 1
|
|||||
Date
|
Time
|
Set Temperature
° C
|
Observed Temperature
° C
|
Recorded by
|
Remark
|
SOP
No. : XXX/SOP/000-00 Format No. : XXX/FRM/000-00
|
Annexure – 2
|
|||||||
Date
|
Name of Sample
|
A. R. No.
|
Batch No.
|
Sample
|
Used by
|
Remark
|
|
Kept
|
Taken
|
||||||
SOP No. : XXX/SOP/000-00
|