1.0 Objective:
The purpose of this SOP is to
provide guidelines for the Disposal of used & waste media in microbiology laboratory.
2.0 Scope:
This SOP covers disposal of used and waste media from
microbiology laboratory.
3.0 Responsibility:
Microbiologist:
Responsible for dispose the media as per guideline and maintain the necessary
record.
QA Officer/QA Manager: Review the records and governing the document.
4.0 Procedure:
4.1 Disinfect the glassware containing
used media and other accessories used for microbiological testing as
follows :
4.1.1 Used media: Pour 5 to 10 ml of 2 % v/v
Phenol or 5 % v/v Dettol solutions to each of the glassware containing used
media for disposal. Use the solutions alternately.
4.1.2 Allow to stand for 30 minutes.
4.2 Sterilize the glassware by autoclaving
15 lbs (at 121° C) for 30 minutes.
4.3 After the pressure in the autoclave is
released, unload the glassware in hot condition and drain the liquid media in
the washing area.
4.4 Wash the glassware .
4.5 After washing the destruction load,
clean the washing area with 5 % v/v Dettol.
4.6 Wash the autoclave with 5 % v/v Dettol
solution and then with tap water thoroughly before using the autoclave again.
4.7 Enter the details in disposal of
microbiological media log book
4.8 CULTURE TUBES:
Disinfect media tubes containing cultures by the above procedure after the freshly sub-cultured is done and
growth observed.
4.9 PIPETTES: Dip used pipettes in a
cylinder or waste bin containing disinfectant. Allow to stand for 30 minutes and
wash the pipettes.