Friday, 16 February 2018

SOP FOR OPERATION OF DO METER.



1.0     Objective:
            To provide written procedure for the operation of DO Meter.
2.0       Scope: 
            This SOP covers operation of DO Meter in Environment Department.
3.0       Responsibility:
Jr. Research Officer, Research Officer: Responsible for operation, calibration and maintenance of the instrument as per procedure.
            QA Officer / QA Manger: Review the records and governing the documents.
3.0       Dissolved Oxygen (DO) Measurements:
3.1       ON/OFF       : This switch is used to switch ON the instrument at the time of use.
                                    It should be switched OFF when not in use.
3.2       SET ZERO   :  This is used for setting 000 on the display when electrode is put
                                     in sodium sulphite solution.    
3.3       Temp. ºC.     :  The knob is kept at a position equivalent to the temp of the      
                                     solution. It compensates for the variation of temperature.    
3.4       CAL              :  The calibration control is used to calibrate the instrument with a
                                     standard solution.
4.0       DO Probe Assembly:
            DO Probe consists of a silver and gold Electrode cells and an electrolyte tube with a ring.
            The following procedure may be followed for assembly of DO Probe.
4.1                Place one membrane on the lower part of the electrolyte tube. Be careful to handle
             its membrane.
4.2                Fix the ring on the electrolyte tube covering the sides of the membrane. Make sure that there are no wrinkles in the membrane & that it does not get puncture.
4.3                Fill potassium chloride solution (7.5%) in the electrolyte tube to its top.
4.4                Insert the electrode in the electrolyte tube and tightly screw it up. Now the probe is ready for use.
5.0                PRECAUTIONS:
5.17        .5% KCL solution acts as a good electrolyte in DO Probe. Presence of any other salt in   
             electrolyte solution may damage the probe.
5.2             The electrolyte tube should not contain any air bubble, if it is present, it will cause      
             fluctuations in the reading
6.0              DO OPERATIONS:
6.1        Set the SET ZERO Knob to the extreme left position and CAL Knob to extreme right
             position                                                                                                                                                    
6.2              Place the DO Probe in 2% sodium sulphite solution allow the display to settle. If it does not read 000set the SET ZERO Knob to bring the display to read 000 by keeping the                  
             temp.ºC knob at the solution temp. CAL Knob should be at extreme right position.          
6.3       Now calibrate the instrument with a known value of DO solution (say distilled water)
6.4       Take 250 ml flask  and fill 2/3 of it with distilled water.
6.5        Stopper the flask and shake it for 30seconds. Remove the stopper back and     
              swirl the water back and repeat the procedure 4 times.
6.6          Measure the temperature of the water and set the Temp. ºC knob at temperature
             of the water in flask.
6.7         Hold DO probe in the flask and agitate the water. Note the reading in the table 1     
            given below for given temperature. If necessary adjust the meter reading with          
            CAL knob to bring the display to read the value as shown in table 1 (eg. At 30ºC,
             display  should read 7.5 in distilled water)
Note:    Please refer to the saturation value of the dissolved oxygen in distilled water as
             given in table-1. It shows the value for dissolved oxygen which will be obtained                     
             when shaking the distilled water at a given temperature in the air.
6.8         Now the instrument is ready to take measurement of DO value of any unknown  
7.0                Caution:       
1)         Since the measurement is based on the diffusion process, the agitation of the solution is must. Hand shaking of the solution may not given accurate readings. Magnetic stirrer is recommended for this process.
2)         Movement of the solution be about 2 ft. /sec for accurate measurement.
TABLE – 1
DO SATURATION VALUES OF DM WATER
ºC
ppm
ºC
ppm
ºC
ppm
ºC
ppm
0
14.3
13
10.3
26
8.1
39
6.6
1
13.9
14
10.1
27
7.9
40
6.5
2
13.6
15
9.9
28
7.8
41
6.4
3
13.2
16
9.8
29
7.7
42
6.3
4
12.9
17
9.6
30
7.5
43
6.2
5
12.5
18
9.4
31
7.4
44
6.1
6
12.2
19
9.2
32
7.3
45
6.0
7
11.9
20
9.0
33
7.2
46
5.9
8
11.6
21
8.9
34
7.1
47
5.8
9
11.3
22
8.7
35
7.0
48
5.7
10
11.1
23
8.4
36
6.9
49
5.6
11
10.8
24
8.3
37
6.8
50
5.5
12
10.6
25
8.2
38
6.7
51
5.4




8.0       HISTORY OF REVISION: 
Revisiom No
Effective Date
Revision Date
Reason for revision





No comments:

Post a Comment

requirement of Q.C in Daman

  We have requirement of Q.C Working area : Daman/Dalwada Exp: Fresher Gender: Male Qualification: Any Graduated Interested Person Call On t...