Thursday, 15 February 2018

SOP for High Volume Air Sample




1.0     OBJECTIVE:
                        The provide written procedure for High Volume Air Sample Model – HVS-1,2,3
2.0       SCOPE:
This Standard Operating Procedure covers operation, Calibration & maintenance of the Instrument.
3.0       RESPONSIBILITY:
                        Supporting Staff/ Jr. Research Officer/ Research Officer
                        QA Officer/ QAM: Governing the Document

4.0       PRE-SAMPLING TASKS:
4.1       SELECTION OF SAMPLING SITE:
            Since sampling is usually effected at 1.5 meter height, raise the Sampler on the stand  or  Table.  Ensure  that filter  is  parallel  to  the  ground. To obtain  a   representative sample, it should not be placed under a tree, near a wall or other obstruction that would prevent free air flow atmosphe  During inclement weather (including high winds), move the entire sampler to  protected location for servicing when practical. While the sampler is removed from the shelter or before the new filter is installed, always remove loose particles from  the inside surfaces of the shelter cover and from the surfaces around the filter                            holder, by wiping them with a clean cloth. Protect the clean installed filter when returning the sampler to the sampling site.

4.2       SELECTION OF FILTER MEDIUM
            Criteria for filter selection  are good collection efficiency, high particles & mass loading  capacity,  low  flow resistance, temperature stability and non- hygroscopicity.
   Glass  micro  fiber  selection  are  particularly suitable for determination of  particulates because of their high retention efficiency combined with low pressur drop, high resistance of blocking and low alkalinity for moisture. Choice of two types of filter i.e. GF/A, and EPM-2000 are available for high volume sampling. Whatmann, GF/A, Glass Fiber filter has been in use for more than 30 years and can be used safely where low trace element analysis is not required. EPM-200 are spectro quality grade glass fiber filters and contain very low levels of inorganic and organic contaminants. They are recommended for use where addition chemical analysis is anticipated. EPM-2000 has been approved by Environmental Protection.Agency  of U.S.A. for  National Air Surveillance Network. It does not contain and binding material and maximum purity is ensured by the use of heat treatment after manufacturing to remove any traces of organic material. Only the filter having  collection efficiency of more than 99 % for particles of 0.3 micron diameters ( as measured by the DOP test ASTN-D2986-71) are to be used the two types of the glass fiber filters   meet this criteria 
Particulate matter that is oily, such as photochemical smog or wood smoke, may block the filter and cause a rapid drop in airflow at a non-uniform rate. Dense fog or high humidity can cause the filter to become too wet and severely reduce the air flow through the filter. Glass-fiber filter filters are comparatively insensitive to  changes in relative humidity, but collected particulates can be hygroscopic. 
4.3       PREPARATION OF FILTER
            Prior to use, each filter should be visually inspected using a light source. Remove loose fibers with a soft brush. Discard or return to the supplier the filters with pinholes and other defects such as tears, creases, or lumps. Always handle the filter papers from its edges and do not crease or fold the filter prior to use.

5.0       IDENTIFICATION, CONDITIONING & WEIGHING:
Assign a serial number to each filter. Stamp this number on two diagonally opposite          corners, stamp on each side of the filter. EPM-2000 comes numbered directly from the
manufacturer. To prevent the filter from getting damaged apply very litter pressure by
pressing it gently.
Filter papers, both blank & containing samples should be conditioned containing active     desiccant for 24 hours before weighing it to minimize errors in the weight; longer periods
Of equilibration will not affect accuracy. The conditioning environment should average     between 20°C and 25°C and not very more than ±3°C with relative humidity (RH) less                        than 50% without varying more than ±5 % RH. A convenient working RH is 40%.
Weight the filters to the nearest milligram preferably up to 0.1 mg and record the weight
and filters identification number, data in data sheet. Ensure keeping a perforated plastic
bottle filled with active silica gel in weighing chamber for minimizing effect of moisture
during weighing.
6.0       OPRATION OF ELECTRONIC CONTROL MODULE:
An explained earlier, all ECMs i.e. EDM1, ECM2 & ECM3 have minimum common basic
Function of Timer/Time  Totaliser. This fuction can be sub-divided into two modes i.e.
DELAY MODE and SAMPLE MODE.
The Delay Mode provides an option to the user to start the sampling operation after a user
Settable delay time. The sample mode refers to the direct operation of sampling with immediate running of the blower.
At power on of the control module, the first message is
During this massge, the micro controller checks the integrity of the internal non-Volatile
memory. If any error if observed, the same is displayed. For details see messages of error
conditions.                  
7.0       DELAY MODE:
            This mode should be used for delayed operation of sampling mode wherein the first step
            is to the delay time.
8.0       Setting of Delay Mode:
            Press DELAY TIME key & the message on display will be as follows
            To set delay time press ENTER Key. With this selection, a blinking cursor would appear
            On the digit of Hours. Use ↑(Increment) ↓(Decrement) to set the desire delay time of        Hours. Every press of ↑ key will increase the hour value by one. Every press of ↓ key will
            Decrement the hour value by one.
ease note that if either ↑/↓ is kept pressed, then the digits will keep on incrementing/
            Decrementing till the key is released. Reset key can be used to make the digit value to       Zero. Once the desired value for Hours is set on the display, press Enter Key, the cursor
            will  shift to the digit of Mts. Again use ↑/↓ keys to set the Mts. Press of Enter key will
            complete the setting of the Delay Time.
9.0       ACTIVATION OF DELAY MODE:
            The delay mode will only get activated when of start key is pressed. The display will show
The colon in lower line will be blinking every second. The time left will be down counted till it becomes zero at which instant, the system will automatically switch over to Sample
            Mode and sampling will start.
            If any power failure occurs during this mode, then all the time data is stored in the memory & the system will start automatically, when power is resumed, with following message for a short time.
After which then earlier message (3) of Time left will be shown. The operation of Delay mode can be stopped with the press of stop key & the display will show message no. 3
            With no colon blinking.
            If power goes off  at this time, then this status is stored & same message will be shown on resumption of power.
Press of Start key will resume the operation of delay mode.
10.0     MODIFICATION OF DELAY TIME:
The modification of set delay time during operation of delay mode is not allowed. It can
be done only after stopping of delay mode with stop key. Press ENTER key, the cursor    will appear on Hrs. digits. Change the values with ↑/↓ keys as explained earlier for Hrs. &
Mts to the new values. The press of ENTER key will modify and store the new values.     Press of start key will resume the operation of Delay mode with new delay time values.
11.0     SAMPLE MODE:
            During this mode, the process of sampling of SPM/REPM will proceed with start of blower. At the end of sampling, the blower will stop. The time of sampling can be set
            on the display.
12.0     SETTING OF SAMPLE TIME:
            To set Sample Time, first press the Sample Time Key. With this, following message will
            be shown with the cursor being off.  
In this message, the second line will be blank, at the start of sampling. To set the Sample
            Time, press ENTER Key. The cursor will appear on the digit of Hrs.
            Subsequent to this, the process of setting of the hours and min. value is similar to Delay
            Set Time, as explained earlier i.e. use ↑ & ↓ key to change the Hrs. digit. Press ENTER Key to move cursor to Minute digits & so on till cursor disappears.
13.0     OPERATION OF SAMPLE MODE:
            At the press of START Key (in case of Delay time is zero) the blower will get ON immediately. The second line of the display will start showing the Sample Run Time with
            the colon blinking every second.
            The above message will also appear when the Time Left becomes zero if delay mode has
            been used. Thus, at the start of this phase, the ‘Run Time’ will be zero and every minute, this Run-Time will be duly incremented by 1 minute. At the end of sampling i.e., when the sampling phase equals the Set-Time, then the sampling phase is terminated and the blower as well as the sampling is stopped. With this following message is shown.
            With this message, 3 long audio beeps are emitted to signify end of sampling.
            During the operation of the sample mode if stop key is pressed any time, the blower power will be put-off and sampling will stop. The Run-Time on the display message 7 will freeze, colon blinking will stop. If power goes off during this mode, then on resumption of
            power, following message will appear,
The sampling can be restarted any time with press of start key. However if power fails during sampling mode, all the relevant time data will be stored in memory and on power
            Resumption, the sampling will start automatically from the point of interruption. The first
            Message for a few seconds will be message 1 for few seconds the following message will
            be shown. This will be followed by sample mode Run message (7)  with last time values will be on shown in display blinking colon.]
14.0     MODIFICATION OF SAMPLE TIME:
            As explained in delay mode, the sample time can also be modified provided the sampling has been stopped with press of stop key. With time frozen on display, press ENTER Key.
            The cursor will appear on digits of Hrs. of set time. Modify this value with ↑/↓ keys. Press
            ENTER Key & the cursor will shift to Mts; set the new value of Mts. At press of ENTER
            Key, the modified sample time value be stored in the memory provided that the new sample time vale is not less than present value of Run-rime as seen on display. However,
            if by mistake the modified sample time value is set less than rum-time, then following message will be shown for short time.
15.0     DISPLAY OF RUN-TIMER AFTER COMPLETION OF SAMPLING RUN:
            You can view this value, by pressing the Sample Time key to display the Sample-Set-Time & Run-Time The message 7 will appear on display with sample time & Run time being equal. Similarly, to observe Delay Time value, press Delay Time Key. The set Delay time & Time left will appear on display. This key can also be used in RUN subsequent to such display values, the above message 8 of run completion is shown again.
16.0     Start of Fresh Sample Run after Completion of Previous Run
            When the message 8 of sample run completion appears on display, then the values of previous Run completion, can be reset by pressing RESET Key or Start Key. With this,
            Following LCD message is shown with the cursor on Y (of Yes)
            Note that, press of Enter key with cursor on yes will reset all run time values to zero. This will also be true for time Left values. However, set value of Delay time & sample time will remain unchanged.
17.0     ADDITIONAL FEATURES OF ECM 2
            As mentioned earlier, model ECM2 suitable for brush type blower has an extra feature of varying the sampling flow rate through keys instead of using a potentiometer. This feather allows precise setting of sampling flow in terms of percentage of speed control of blower. To activate this mode, press Misc. Key following message will appear
            Press ENTER key and the cursor will appear on numerical values. Use ↑/↓ keys to increase  or decrease the speed. Press ENTER key again to come out of this mode.
            Please note that there is no one to one relationship between percentage value of speed
            control on display & the flow value as determined by measurement of DP on manometer and corresponding flow value from the calibration graph.
18.0     ADDITIONAL FEATURES OF ECM 3:
            CONSTANT FLOW CONTROL DURING SAMPLING
            In addition to the timer totalizer functions, the ECM 3 has additional feature of maintaining the sampling flow constant irrespective of dust load and mains supply voltage
            variation. This is achived by increasing or decreasing  the sampling flow rate through the blower speed control.
            *The Sampling flow is set by moving the sensor block to the required set position on the U-Tube manometer.
            Procedure for Setting the Required Flow:
            As explained earlier, the sampling flow rate should be set at 1.13 M3/Min. for SPM. This is achieved by first finding the equivalent differential pressure from the calibration graph & then moving the sensor block to the required position. The detailed procedure is given bellow:
            1.         Decide the Sampling flow
2.         Refer to calibration graph of flow v/s differential pressure in CM WC on U-Tube. Read the DP value on X-Scale by drawing a horizontal line from the Y-axis for the set flow value to the curve. Read corresponding value of DP on X-Scale by drawing a vertical line from the intersection on the curve & horizontal line.
3.         Ensure that water level in manometer is filled upto 0-0 level & there are no bubbles in water.
4.         Divide the value of DP by 2 and move the pointer sensor block upward from zero level to half of DP value.
5.         Once the blower starts the water level in manometer right limb will go up while the left limb it will go down. Once it reaches the sensor block its level will be detected by the sensor & this information will be sent to the micro-controller. This position
            of water level will be maintained by the micro-controller. The difference between
            the water levels on the two limbs will be corresponding DP Value of the required set flow.
19.0     USE OF MISC. KEY:
This key has two purposes. Before the sampling start, this key can be used to effect preliminary check for the functioning of the two sensors in the manometer sensor block. During the Run Mode also, this key can be used to check the flow status from functioning of these two sources. As can be noted that this flow status is derived from the water level in  the glass manometer with respect to the Top and Bottom Sensors of the manometer Sensor Block. Thus when the water level is above or below both the sensors, then the flow
is said to be above or below the flow set value. It is only when the water level is in between the top and bottom sensors, that the flow is declared to be at set flow:





20.0     USE OF MISC. KEY BEFORE THE RUN MODE:
Press of this key before sampling is used to check proper functioning of the sensors. Following message is shown when this key pressure
During this test, the blower is started the flow (and hence the water level) is expected to increase sufficiently high, so that both the top and the bottom sensor can be tested properly
for pressure of water.
Press the ‘START’ key. With this, at first following message is shown and the blower also
Get started.
With blower on, as the flow increases and the resultant ΔP across the blower ‘Exhaust-Orifice’ causes the water level in the manometer to rise at first above the bottom sensor. With this, the following single line message is shown:
When the flow develops sufficiently high so as to cause the water level rinse above the top
sensor as well, then the Top sensor O.K! is displayed on the second line of above message.
Thus, when both sensors are declared functionally OK as shown in the above message, then the blower is put-off. This completes the manometer flow test.
21.0     USE OF MISC. KEY DURING THE RUN MODE:
During run mode, when this key is pressed, then one of the following three messages is
shown appropriately, as per the water level in the glass manometer with respect to the two
sensors in the manometer sensor block. These three messages are as follows:
This will be shown when water level remains in-between the two sensors of the manometer sensor bock                                                                                                    
                                                OR
This will be shown when water level remains below both the bottom and top sensors of the manometer sensor block.
                                                OR
This will be shown when water level remains above both the bottom and top sensor of the manometer sensor block.
Note that the Control Module electronics always monitor the water level in the glass manometer and whenever the level does not remain in between the two sensors, then the
blower speed is adjusted in appropriate direction so that the condition corresponding to
Manometer at Set-Flow! Is maintained.
22.0     LCD MESSAGE DURING RUN-SAMPLE-MODE
            At the start of this phase, the ‘Control Module’ starts the blower motor to a predetermined
            starting speed. With this, the blower motor is started and following is shown:
Note that at the start of this Run-Sample-Time, the second line in this message is kept black for a while.
Note that the manometer Sensor Test is effected only at the start of a Fresh Sample-Run
(i.e. for Run Time beginning from zero). This is to ensure satisfactory operation of both the sensors of the Manometer Block, right from the start of the Run-Mode, this Test is not
effected.
After the blower start, HVS flow increases slowly and the water level in the glass manometer goes on rising above its initial level. Once, it reaches above bottom sensor, then following message is inserted in the second line of above message:
After the above message, the flow increases further and the water level rinses even above
the top sensor. With this following message is shown in the second line:
In this way, the software tests the proper functioning of both the sensors.

23.0     CONSTANT FLOW CONTROL ACTION DURING SAMPLE-RUN-MODE
After the above test of Manometer Sensors, the ‘Control Module’ duly adjusts the blower
speed in such a way that the water level always remains in between the Top and the Bottom Sensor of the Manometer Block, i.e. the blower flow is kept constant at the preset value. In Sample Run Mode, whenever the water level deviates from this situation, then the blower speed is duly increased / decreased so as to again restore the water level in between the two sensors.
24.0     ERROR CONDITIONS FOR MANOMETER SENSOR BLCOK WITH MISC. KEY PRESS :
A.    In this ‘Mano. Flow Test’, at first, the “START” key is expected to be pressed. Thus, at the press of “START” key (but before the blower start), any of the following Error situations may be encountered: After pressing the START key, if the water level in the Manometer is not below both the manometer sensors, then following two messages are shown one after another:
B.     After pressing the START Key, if the water level in the Manometer is below Top
Sensor but above the Bot. Sensor, then following two messages are shown one after another.
C.    After pressing the START Key, if the Top Sensor signal is found to be corresponding to water level above this sensor but Bottom Sensor signal is not found to be active (i.e.
Bot. Sensor signal does not correspond to water level above it), then at first following
Three messages are shown one after another:
In this Error situation, check if the Manometer Block position is properly placed and not over tightened nor kept loose.
D.    At the end of Mano.Flow Test, if the water level fails to reach above the two sensors, then following two messages are shown continuously one after another
25.0     ERRORS AT POWER-ON AS REGARDS ‘CORRUPTION OF NONVOLATILE MEMORY’:
In case such error has been encountered at power-on, then following two messages are shown with initialization action of the entire Nonvolatile memory starting from 00 onwards.
With the above initialization action, all the data as regards Set and Run time values of the
Delay-Phase and Sample-Phase gets initialized to default values.
26.0     METHOD FOR SAMPLING OF GASEOUS POLLUTANTS:
As explained earlier, the Sampler has provision to collect four gaseous pollutants simultaneously while sampling for particulate matter. In case the gaseous pollutants are to be collected, you may follow the methods that are approved by the Central Pollution Control Board or Bureau of Indian Standards. These methods for SO2 & NO2 have been
Described in detail in Appendix II. For, further information on sampling of other pollutants i.. H2S, NH2, Cl2 etc., please check with Central or State Pollution Control Board or BIS.
The method involves filling each of the impinger tube with a specific adsorbing solution and then passing known volume of ambient air through the solution at an appropriate flow rate. The concentration of the pollutants can then be determined in the lab using specific reagents and spectrophotometer or colorimeter as per detailed procedure given in each method.
In Practice, take out an impinger tube and fill it with the adsorbing solution for SO2 take out 2ng impinger & fill it with specific reagent for absorption of NO2.
Put crushed ice or cold water in the impinger box from the opening impinger tube. This helps in better absorption of pollutants & also reduces vaporation of reagents when ambient temperature is high. Follow these steps.
i.        At first set a required Sampling Time at the Control Unit. Also put the filter paper on the filter holder.
ii.                  Now, put-on the HVS equipment. With the Blower ON, set the desired sampling flow by controlling the respective needle valve of the manifold after rotameter to the centre tube of the respective impinger & check the flow rate (Refer the Flow Schematic Diagram at the end of the Manual) reading.
At the end of the Sampling, again connect rotameter to the central tube of each impinger note down this and note down the flow rate. Take the impinger tubes to the laboratory for analysis. Drain out the water at the Impinger tube Housing by removing the drain plug. Fix it back after all water is drain out.
Take average of the initial & final flow, compute the average flow and compute volume of sampled air by multiplying sampling time in mts to the average flow i.t.
                                                                  Initials Flow + Final Flow
Volume of sampled gaseous pollutant = -------------------------------- (LPM) X Time (mts.)
                                                                                        2
Please note that gaseous pollutants are normally collected for 4 hrs while particulate sampling will be for 24 hrs. There fore it is important to note down the sample time & final rate to achieve better accuracy.


DO NOT KEEP THE ROTAMETER TUBE CONNECTED TO THE CENTER TUBE OF ANY IMPINGER WHILE SAMPLING GASEOUS POLLUTANTS.

27.0     POST SAMPLING TASKS:
            REMOVAL OF FILTER PAPER
            The following procedure should be used to remove the exposed filter
            1.         Remove the pressure plate and lift the exposed filter from the supporting screen by
                        grasping it gently at the ends. Under no circumstances one should try to remove filter paper with blower on.
            2.         Fold the filter lengthwise at the middle with the exposed side in, if the collected  sample is not centered on the filter (i.e. the unexposed border is not uniform     around the filter), fold so that only the deposit touches the deposit. An  improperly folded filter where smudge marks from the deposit extend across the borders can  reduce the value of the sample if the analysis for which the sample was collected needs to be divided into equal portions.
                        
            3.         Check the filter for signs of air leakage. Leakage may result from a worn pressure plate gasket or from an improperly installed gasket. If leakage signs are observed,
                        void the sample, determine the cause, and take corrective actions.
            4.         Inspect visually the gasket face to see if glass fibers from the filter are being left behind due to over tightening of the face plate wing nuts and to consequent cutting  of the filter along the gasket interface. One can use small amount of talcum powder                  to provide a thin layer on the gasket.
            5.         Check the exposed filter for physical damage that may have occurred during or  after sampling. Physical damage after sampling would not invalidate the sample if
                        all pieces of the filter were put in the folder; however, sample losses due to  leakages during the sampling period or losses of loose particulates after sampling
                        (e.g. loss when folding the filter) would invalidate the sample, so mark such samples void before forwarding them to the laboratory.
            6.         Check the appearance of the particulates. Any change from normal colour for example, may indicate new emission sources or construction activity in the area.
            
Note any change on the filter folder along with obvious reasons for the change.

28.0     FILTER LITERATURE AND INSPECTION:
            A.        Remove the filter folder from its shipping envelope.
            B.        Record the filter number on the hi-vol. field data laboratory date log.
            C.        Examine the shipping envelope. If sample material has been dislodged from the  filter with a soft camel’s-hair brush.
            D.        Examine the filter. If insects are embedded in the sample deposit, remove them with Teflon-tipped tweezers, but disturb as little of the sample deposite as possible  If more than 10 insects are observed, refer the sample to the supervisor for a                                    decision to accept or reject it.

29.0     FILTER CINDITUONING:
            The exposed filters should be conditioned in a desiccator containing active desiccant as    was done for 24 hours; or move before sampling. One can use an equilibration chamber
            with a saturated chemical solution to get an RH of less than 50% at 20°C. The chamber
            is recommended in lieu of a controlled-temperature/humidity weighing room. And Air
            Conditioned room may be used for equilibration if an RH of less than 50% that is constant
            within +/- 5 % and an air-temperature that is constant within ±3°C are maintained while    the filters are equilibrating. A convenient working RH is 40 % keep hygrometer in the     room.

30.0     GRAVIMETRIC ANALYSIS:
            Weight the exposed filter within 30 sec. after removing it from the equilibration chamber to the nearest milligram (mg) preferably up to 0.1 mg on the analytical balance.
            Weigh the filter in the conditioning environment if possible. If not, be sure that the            analytical balance is as close as possible to the conditioning chamber where it is relatively       free of air currents and where it is at or near the temperature of the chamber.
            Record the weight in the laboratory data log and on the hi-vol. field data form along with
            the filter paper before sampling.
31.0     EVALUTING THE VOLUME OF THE AIR SAMPLED:
            For Simple evaluation of volume of air sampled, follow these steps
            a.         Note down the differential pressure DP of the water level at manometer at the start of sampling and find the corresponding flow from calibration graph. Let it be                                      called Fs.
            b.         Similarly, note down the DP of manometer & the corresponding flow at the end of  sampling. Let it be called Fe.
            c.         Find the average flow as follow
                                                                Fs + Fe
                                                Favg  =  ------------    Cu.mt/min
                                                                      2
            d.         Multiply the Favg with time of sampling in Minutes to get volume V of sampled                            air.                        V  =  Favg  X  Time ( im mt.)

32.0     For Systematic evaluation of the air volume sampled follows these steps:
            1.         Note down the differential pressure with the manometer after every hour and fixed  the corresponding dlow in cubic meter per min. from the calibration graph                                           provided along with the equipment.
            2.         At the end of Sampling, also note down the differential pressure value and the flow value.
            3.         For every hour after the Sampling, start, find out the average flow by adding the flow values at the start and at the end of the hour and divide the resultant addition by two.
            4.         For every hour, multiply the average flow for this particular hour by 60 minutes to
                        find out volume in cubic meters in this hour.
            Evaluate the total volume by adding all the volume in cubic meters for every hour.
33.0     CALCULATION OF SPM CONCENTRATION:
            The weight of the particulate matter is then divided by the total volume of the sampled air
            to determine the dust concentration in microgram per cubic meter.
                        For calculating the SPM concentration use the following equation
                                                                                                   (Wf  -  Wi)
                                    S.P.M. Concentration (µg/m3)   =    ----------------------- X  106
                                                                                                           V
                        Where  V    =   Volume of air sampled in cubic meters
                                    Wf   =   Final weight of exposed filter in gms.
                                    Wi   =   Initial weight of filter in gms.
                        Record all original calculations in the data log book.
            Of DP on manometer and corresponding flow value from the calibration graph.

34.0     History of Revision:
Revision No.
Effective Date
Revision details
Reason for revision


No comments:

Post a Comment

requirement of Q.C in Daman

  We have requirement of Q.C Working area : Daman/Dalwada Exp: Fresher Gender: Male Qualification: Any Graduated Interested Person Call On t...