1.0 Objective:
The objective of
this standard Operating Procedure is to lay down the procedure for Operation and Calibration of Nephelometer.
2.0 Scope:
This SOP covers
operating and calibration of Nephelometer
3.0 Responsibility:
Jr. Research Officer, Research
Officer: Responsible for operation, calibration and maintenance of the instrument as
per procedure.
Head of Department: Responsible for
calibration and maintenance, timely as per schedule.
QA Officer/QA Manager: Review the records and governing the document.
4.0 Procedure:
4.1 Switch ON the instrument (back side switch).
4.2 Open the lid of the sample
compartment & insert a test tube filled with the distilled water into the sample
compartment then close the lid.
4.3 Rotate the switch of
“set 0” to get “0” on LCD.
4.4 Open the lid,
replace the test tube of distilled water by Formazine Std
.Solution, and close the lid.
4.5 Rotate the switch of
“Set 100” to get “100” on LCD.
4.6 Take the reading of
sample in same way after setting standard.
4.7 Switch of the
instrument.
Note: Wash the test tube
with distilled water after taking result.
5.0
Calibration Procedure:
Solution 1:- Weigh accurately 5.0 g of Hydrazine Sulphate in 500 ml
volumetric flask and add distilled
water to make up to the mark. Leave the mixture to stand for 4 hours.
Solution 2:- Weigh accurately 50.0 g of Hexam-Ethlenetetramine in 500ml volumetric flask and add make up the volume up to the mark with distilled water.
Mix the equal volume of solution 1 and 2 form Formazine Turbidity Concentrate Allow to stand in a closed container at 25º C to 30º C for 48 hrs to produce insoluble while turbidity corresponding to 4,000 NTU (This solution is stable for 2-3 Months).
Formazine Standard -100 NTU :- Dilute 25 ml of the Formazine Turbidity Concentrate to one liter with turbidity free distilled water to obtain 100 NTU Formazine Standard for setting 100 of the instrument. (This Solution should be prepared weekly)
6.0
Routine Maintenance:
6.1
Clean the instrument with dry
cloth.
6.2
Clean sample tube with methanol
and than with water.
6.3
Clean the sample holder with
dry cloth.
7.0 Documentation:
Annexure – 1 - Usage log book . XXX/FRM/000
Annexure – 2 - Calibration record. XXX/FRM/000
8.0 History of Revision:
Revision No.
|
Effective
Date
|
Revision
details
|
Reason for
revision
|
Annexure - 1
|
||||
Date
|
Booking No.
|
Name of sample
|
Remark
|
Sign
|
SOP No. : XXX/SOP/000-00 Format No. : XXX/FRM/000-00
|
Annexure - 2
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
CALIBRATION DATA OF NEPHELOMETER
|
Page No.
01 of 01
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Instrument: Nephelometer
ID# :
Make : Model :
Month:
Frequency: Daily
Calibrated by:
Checked by:
SOP No. : XXX/SOP/000-00 Format No. : XXX/FRM/000-00
|