PRINCIPLE:
The
Cadmium content of the sample is determined by Pre-Concentrate aspirating the
sample into the flame of an Atomic Absorption Spectrophotometer. The Absorbance
is measured at 228.8 nm using a Cadmium Hollow Cathode Lamp.
Limit of Detection: 0.01 mg/L
REAGENTS
- Hydrochloric acid – Concentrated
- Nitric acid : Concentrated
- Nitric acid : (1:499)
- Cadmium stock Solution : Merck 1000 mg/L NIST Certified Reference Material or Dissolve 1 g of pure cadmium metal in minimum quantity of concentrated nitric acid and dilute to 1 liter with distilled water (1 ml = 1 mg of Cd)
PROCEDURE
Preparation
of standard solution: Prepare a reagent blank and a
series of 100 ml standards containing 0.1, 0.2, 0.3, 0.4 and 0.5 mg/L of
cadmium by diluting a suitable volume of the standard solution with dilute
nitric acid.
Sample
preparation: Transfer 1000 ml well-mixed sample to
the evaporation dish and add 3 ml nitric acid. Heat and evaporate to 20 ml (do
not boil). Cool and add 3 ml nitric acid and heat until digestion is complete.
Add 2 ml HCl (1:1 v/v) and heat gently to dissolve the residue. Wash the
evaporation dish for 3 to 4 times with water and filter. Wash the filter and
discard. Dilute the filtrate & make up 100 ml volume with water. (10 times
concentrate the sample)
Analysis: Operate the instrument as per SOP. Aspirate nitric acid (1:499) as
blank. Then aspirate series of standards and then aspirate the sample solution.
Read the concentration in mg/L of zinc directly.
(Apply the
dilution factor 0.1 because the sample is concentrate 10 times)