Friday, 21 September 2018

SOP FOR OPERATION AND CALIBRATION OF GAS CHROMATOGRAPHY (THERMO-TRACE)


1.0         Objective:
              The objective of this Standard Operating Procedure is to lay down the procedure for
              operation and calibration of Gas chromatography (THERMO-TRACE)
2.0         Scope:
             This SOP covers operation and calibration  procedure of  Gas chromatography
             (THERMO-TRACE).
3.0          Responsibility:
             Jr. Research Officer, Research Officer: Responsible for operation and maintenance of
               the instrument as per procedure.
Head of Department: Responsible for maintenance, timely as per   schedule.
QA Officer/QA Manager: Review the records and governing the document
4.0         Procedure
4.1          Ensure that the instrument is clean and free from dust. Wipe all the traces of
               solvent/water/moisture by dry cloth.
4.2          Properly install suitable capillary column in GC oven injector & selected detector.
4.3                    Open the Gas pressure valve from the distribution panel. (N2 cylinder for N2 set 60 psi for N2).
4.4         Check the backpressure of column.
4.5         Switch on the mains power, stabilizer, GC power & computer power.
4.6                    Double click on the Chrom-card icon from the desktop, chrom-card window display on
               the screen, double click on Konark icon, select the method from edit G.C parameters
icon, then.   Apply the chromatographic condition and enter ok open run signal. After
               completion of the run, open the instrument in offline, then go to reprocessing menu and
               select file and click on ok button, integrate peak, after integration enter the component
               name then click on ok  ,go to edit method than go to in report parameters and click on ok
               to view the print layout print chromatogram by clicking printer icon.
4.7                    Condition the column at 280°C for 10 min. Enter the programming condition of the sample through micro syringe, when G.C shows ready signal.
4.8                    Cool the G.C after completion of the analysis & switch it off. Shut down the computer,
               mains power supply. Close the gas supply from the distribution panel & from cylinder.
5.0         Calibration procedure:
5.1         Calibration of Flow rate
5.1.1      Open the column oven compartment.
5.1.2      Remove the column if attached to the injector port –1 (to be calibrated).
5.1.3     Connect soap film flow meter (side arm) and injector port outlet with the help of Teflon or rubber tube.
5.1.4          Fill the pipette bulb partially with a soap solution and attach to the bottom of the flow
              meter.
5.1.5          Open the knob of carrier gas (N2) and set up proper pressure (i.e. 500 kpa or 5 Kg/cm2) 
   in carrier gas pressure controller.
5.1.6      Check for leaks at each and every point of attachment using soap solution.
5.1.7      Open the knob of carrier gas flow controller & allow carrier gas to flow through
              corresponding digital flow control.
5.1.8     Adjust the carrier gas flow rate with flow control (30 ml/min).

5.1.9     Gentle squeeze the bulb to force a soap film up into the gas stream. Start the stopwatch as
             soon as the film reaches to zero ml mark. Stop the watch when the film reaches to 30 ml
             marking. Note down the time require to reach the film from 0 to 30 ml mark. Calculate
the  flow rate using the following formula.
              Flow rate ml/min  = 60 X 30 (Dist. Between two points 30 ml)
                                                         Time taken in seconds
5.1.10      Similarly calibrate the flow rate after the time interval of 20 min, 40 min, & 60min and
 find out the difference between the readings.
5.1.11      Acceptance criteria: observed flow rate of the equipment should be within ±2ml/min of
 set flow rate.
5.2       Calibration of FPD        
5.2.1    Detector precision and consistency of relative retention time:
5.2.1.1  Preparation of pesticide standard (0.2ppm):
   Take 100µl of standard solution (100ppm) of Methyl parathion in 10 ml volumetric
   flask containing 5ml acetone and dilute to mark with same solvent. Take 2 ml of this
   solution in 10 ml volumetric flask and dilute to mark with acetone.
5.2.1.2  Inject 2.0µl of pesticides standard 5 times and observe area and RT.
5.2.1.3  ACCEPTANCE CRITERIA:
  The deviation of RT ± 0.2 min
  The deviation of area ± 10%
5.2.2     Detector Linearity:
5.2.2.1  Prepare the five different concentration solution as follow to check the detector linearity.
             1.        Preparation of pesticide standard 1 (0.025 ppm):
   Take 100µl of standard solution (100 ppm) of Methyl parathion in 10 ml volumetric flask  
   containing 5ml acetone and dilute to mark with same solvent. Take 1ml of this solution in
   10ml volumetric flask and dilute to mark with acetone. Take 5ml of this solution in 10ml
    volumetric flask and dilute to mark with acetone. Take further 5ml in 10ml volumetric
   flask and dilute to mark with acetone.
               2.      Preparation of pesticide standard 2 (0.05 ppm):
    Take 100µl of standard solution (100 ppm) of Methyl parathion in 10 ml volumetric flask  
    containing 5ml acetone and dilute to mark with same solvent. Take 1ml of this solution
    in 10ml volumetric flask and dilute to mark with acetone. Take further 5ml in 10ml
              volumetric flask and dilute to mark with acetone.     
             3.        Preparation of pesticide standard 3 (0.1 ppm):
    Take 100µl of standard solution (100 ppm) of Methyl parathion in 10 ml volumetric flask 
    containing 5ml acetone and dilute to mark with same solvent. Take 1ml of this solution
     in 10ml volumetric flask and dilute to mark with acetone.
             4.        Preparation of pesticide standard 4 (0.2 ppm):
    Take 100µl of standard solution (100 ppm) of Methyl parathion in 10 ml volumetric flask 
    containing 5ml acetone and dilute to mark with same solvent. Take 1ml of this solution
   in 5ml volumetric flask and dilute to mark with acetone.
               5.        Preparation of pesticide standard 5 (0.4 ppm):
              Take 100µl of standard solution (100 ppm) of Methyl parathion in 10 ml volumetric flask 
              containing 5ml acetone and dilute to mark with same solvent. Take 2ml of this solution
  in 5ml volumetric flask and dilute to mark with acetone
5.2.2.2  Inject 3 replicates injection of the above solution and observes area.
5.2.2.3  Plot the graph of linearity and calculate the correlation coefficient.
               Acceptance criteria:
Correlation coefficient minimum 0.99
5.2.2.4    CHROMATOGRAPHIC CONDITION:
               Name of the instrument                          : GC [Thermo Trace ultra]

               Column                                                    : CPSIL-8CB [30m X 0.25 mm ID X 0.25µm] 
                                                                                                       @ 15°C/min  
               Oven Temp.                                             : 130°C (3min) 270°C (10 min)
               Injector Temp.                                         : 250°C
               Detector Temp.                                        : 300°C
               Carrier Gas                                               : N2 [2 ml] H2 (115ml) AIR (120ml
               Injection Volume                                     : 2.0µl
               Split Ratio                                                : Split less             
5.3       Calibration of NPD
5.3.1    Detector Precision and consistency of relative retention time: -
5.3.1.1 Preparation of pesticide standard (0.2ppm):
Take 100 mL of standard solution (100 ppm) of Chlorpyrifos in 10 ml volumetric flask containing 5ml acetone and dilute to mark with same solvent. Take 2 ml of this solution in 10 ml volumetric flask and dilute to mark with acetone.
5.3.1.2 Inject 2.0µL of above pesticide standard 5 times and observe area and RT.
5.3.1.4 ACCEPTANCE CRITERIA:
The deviation for RT ± 0.2 min and deviation for area ± 10%.
5.3.2    Detector Linearity: -
Prepared the five different concentrate solution as follow to check the Detector Linearity
5.3.2.1 Preparation of pesticide standard 1 (0.025ppm): Take 100 mL of standard solution (100 ppm) of Chlorpyrifos in 10 ml volumetric flask containing 5ml acetone and dilute to mark with same solvent. Take 1 ml of this solution in 10 ml volumetric flask and dilute to mark with acetone. Take 5 ml of this solution in 10 ml of volumetric flask and dilute to mark with hexane. Take further 5 ml in 10 ml volumetric flask and dilute to mark with acetone.
5.3.2.2 Preparation of pesticide standard 2 (0.050ppm): Take 100 mL of standard solution (100 ppm) of Chlorpyrifos in 10 ml volumetric flask containing 5ml acetone and dilute to mark with same solvent. Take 1 ml of this solution in 10 ml volumetric flask and dilute to mark with acetone. Take further 5 ml in 10 ml of volumetric flask and dilute to mark with acetone.

5.3.2.3 Preparation of pesticide standard 3 (0.1ppm): Take 100 mL of standard solution (100 ppm) of Chlorpyrifos in 10 ml volumetric flask containing 5ml acetone and dilute to mark with same solvent. Take 1 ml of this solution in 10 ml volumetric flask and dilute to mark with acetone.
5.3.2.4 Preparation of pesticide standard 4 (0.2ppm): Take 100 mL of standard solution (100 ppm) of Chlorpyrifos in 10 ml volumetric flask containing 5ml acetone and dilute to mark with same solvent. Take 2 ml of this solution in 10 ml volumetric flask and dilute to mark with acetone.
5.3.2.5 Preparation of pesticide standard 5 (0.4ppm): Take 100 mL of standard solution (100 ppm) of Chlorpyrifos in 10 ml volumetric flask containing 5ml acetone and dilute to mark with same solvent. Take 2 ml of this solution in 5 ml volumetric flask and dilute to mark with acetone.
5.3.2.6 Inject 3 replicates injection of the above solution and observe Area
5.3.2.7 Plot the graph of linearity and calculate the correlation-coefficient
5.3.2.8 CHROMATOGRAPHIC CONDITION
Name of Instrument                          : GC [THERMO]
Column                                               : CPSIL-8CB(30.0Mx0.25 X 0.25mm)
Oven Temp.                                        : 130°C (3minhold)@5°C/min-->270°C(10min hold)
Detector                                              : 300 °C(NPD)
Injector                                                : 250 °C
Carrier Gas                                          : N2 [2.0ml/min ]
Injection volume                                 : 2 mL
Split Ratio                                           : split less
5.3.3    Acceptance criteria: -
Correlation-coefficient Minimum 0.99
5.4      Calibration of Oven:
            Calibration of oven to be done by calibration department.
5.5        Frequency of calibration: Quarterly
6.0        Routine Maintenance:
6.1                Clean the all, out side part of instrument.
6.2                Clean the injector and injector liner. Change the glass wool of injector liner. Change the septa of injector port if required.
6.3                Check the all connection of gases for leakage.

No comments:

Post a Comment

requirement of Q.C in Daman

  We have requirement of Q.C Working area : Daman/Dalwada Exp: Fresher Gender: Male Qualification: Any Graduated Interested Person Call On t...