Thursday, 10 May 2018

SOP OF Determination Of NOISE


Principle:       
To describe the general test method for measurement of noise.


Scope:              
This work instruction is applicable for measurement of noise at various locations.


Procedure:                                                
INSTRUMENT: Environmental noise pollution monitor
                        

Test method:                                                

Switch on the instrument by DC switch.                                 
    
Check the instrument for 94 dB(A) for its normal operation before taking for measurement

Place the monitor in such a way that the sensor of the monitor is facing the noise.

Keep the setting of four switches as under                           
                   ON / OFF        -  ON (DC) 
                   Range                   50 - 100
                   Selector switch -  A
                   Hold max          -  F

Equipment directly indicates the noise level.                        

Record the observation.


No comments:

Post a Comment

requirement of Q.C in Daman

  We have requirement of Q.C Working area : Daman/Dalwada Exp: Fresher Gender: Male Qualification: Any Graduated Interested Person Call On t...