1.0 Objective:
To provide written procedure for operation and calibration of Antibiotic Zone Reader
8.0 History of Revision:
To provide written procedure for operation and calibration of Antibiotic Zone Reader
2.0 Scope:
This SOP covers the operation and calibration of Antibiotic
Zone Reader.
This SOP covers the operation and calibration of Antibiotic
Zone Reader.
3.0 Responsibility:
Microbiologist: Responsible for operation, calibration and
maintenance of the instrument. Maintain the required record.
QA Officer/QAM: Review the records and governing the
document.
Microbiologist: Responsible for operation, calibration and
maintenance of the instrument. Maintain the required record.
QA Officer/QAM: Review the records and governing the
document.
4.0 Procedure:
4.1 Ensure that the instrument is visibly clean and free from dust.
4.2 Switch ‘ON’ the mains switch.
4.3 Turn the left side drum to its maximum
clockwise position, so that Zero Marking on the scale of the right side drum
coincides with the Zero marking on the Vernier scale.
4.4 Keep the Petri-dish on the Aluminum
plate and rotate it manually; so that one end of the Zone image touches the Mark Line
on the Prism.
4.5 Now turn the left-side drum slowly, in
anti-clockwise position, till the other end of the Zone image touches the Mark Line
on the Prism.
4.6 Note the reading directly on the Right
side drum scale. The scale is marked from 0 to 35 mm with each division of 0.2
mm and the Vernier scale is marked with each division of 0.02 mm.
5.0 Calibration:
5.1 Operate
the instrument as per above procedure.
5.2 Put
the standard coin in the appropriate place on antibiotic zone reader.
5.3 Measure
the zone of standard coin and compare with the standard coin result and
observed result.
5.4 Record the result in the calibration
format as per given in Annexure - 1.
6.0
Routine maintenance:
6.1
Clean the instrument with dry
cloth.
6.2
Clean glass & lance with
dry cloth or tissue paper.
7.0 Documentation:
7.1 Annexure – 1 Antibiotic Zone Reader calibration
format XXX/FRM/000-00
7.2
Annexure – 2 Antibiotic
Zone Reader logbook XXX/FRM/000-00
8.0 History of Revision:
Revision No.
|
Effective
Date
|
Revision
details
|
Reason for
revision
|
Annexure
- 1
|
||||||
Calibration Record of
Antibiotic Zone Reader
|
||||||
Calibration Date
|
:
|
Next Due Date for calibration :
|
||||
Instrument ID No.
|
:
|
Model No. :
|
||||
Make
|
:
|
|||||
Sr.
No.
|
Standard
Coin Diameter (In mm)
|
Observed
Diameter
(In mm)
|
Remark
|
|||
1
|
||||||
2
|
||||||
Calibrated by:
Checked by:
Date:
Date:
|
||||||
SOP No. : XXX/SOP/000-00 Format No. : XXX/FRM/000-00
|
Annexure – 2
|
|||||
Date
|
Name of Sample
|
A. R. No.
|
Batch No.
|
Used by
|
Remark
|
SOP No. : XXX/SOP/000-00 Format No. : XXX/FRM/000-00
|